CONTENTS I. UPLC INTACT MASS ANALYSIS APPLICATION KIT COMPONENTS II. INSTALLING THE PRE-COLUMN TUBING INTO MASSPREP MICRO DESALTING COLUMNII
II. INSTALLING THE PRE-COLUMN TUBING INTO MASSPREP MICRO DESALTING COLUMNRequired materials: 1/4-inch open-end wrench5/16-inch open-end wrenchF
Figure 4. Fluidic configuration for LC-MS (UPLC-LCT Premier) analysis. A post-column salt diversion valve (top-left corner of the LCT Premier) can be
Figure 5. Fluidic configuration for LC-MS analysis using mass spectrometers with no integrated salt divert valve.Figure 6. Column placement in the co
2. Plug the terminal block into terminal block I/O (terminal block I/O is located on the back panel of the external divert valve).3. Connect the oth
VI. SUGGESTED METHOD FOR MASS ANALYSIS USING UPLC INTACT MASS ANALYSIS APPLICATION KITMass Spectrometry has become a powerful tool for therapeu
Figure 8. Total ion chromatograms (TICs) from Waters ACQUITY UPLC-LCT Premier ESI-TOF MS analyses of an intact IgG1, and pre and post blank runs. Col
MS ConditionsThe MS parameters listed below can vary depending upon the physical characteristics of the analyzed proteins, flow rate, and MS probe pos