CONTENTS
I. UPLC INTACT MASS ANALYSIS APPLICATION
KIT COMPONENTS
II. INSTALLING THE PRE-COLUMN TUBING INTO
MASSPREP MICRO DESALTING COLUMN
III. CONNECTING THE COLUMN TO LCT
PREMIER MS
IV. CONNECTING THE COLUMN TO Q-Tof
PREMIER MS, AND SYNAPT HDMS
V. CONNECTING THE COLUMN TO WATERS SQ
DETECTOR (SQD)
IV. SUGGESTED METHOD FOR MASS ANALYSIS
USING UPLC INTACT MASS ANALYSIS
APPLICATION KIT
a. Intact Antibody Analysis
b. Reduced Antibody Analysis
VII. TIPS FOR SUCCESS
UPLC Intact MASS Analysis Application Kit
(Part Number: 176001519)
I. UPLC INTACT MASS ANALYSIS APPLICATION
KIT COMPONENTS
Description Qty Part Number
MassPREP Micro Desalting Column 1 186004032
Desalting Cartridge Manual Assembly,
Cap/Tube/Column
1 430001988
UPLC Intact Mass Analysis Solution 1 715001664
Intact Mass Tubing Kit. Kit contains: 1 205000548
Peek Compression Screw, Long, 1032 4 410002276
Peek Tubing, 0.062 OD x .004 ID x 8.50 L 1 430001991
Peek Tubing, 0.062 OD x .004 ID x 15.04 L 1 430001992
Peek Tubing, 0.062 OD x .004 ID x 22.20 L 1 430001993
[ CARE AND USE MANUAL ]
P/N: 430001988
P/N: 186004032
P/N: 410002276
P/N: 430001991
P/N: 430001992
P/N: 430001993
Figure 1. UPLC
®
Intact Mass Analysis Application Kit Components.
Page view 0
1 2 3 4 5 6 7 8

Summary of Contents

Page 1 - [ CARE AND USE MANUAL ]

CONTENTS I. UPLC INTACT MASS ANALYSIS APPLICATION KIT COMPONENTS II. INSTALLING THE PRE-COLUMN TUBING INTO MASSPREP MICRO DESALTING COLUMNII

Page 2

II. INSTALLING THE PRE-COLUMN TUBING INTO MASSPREP MICRO DESALTING COLUMNRequired materials: 1/4-inch open-end wrench5/16-inch open-end wrenchF

Page 3

Figure 4. Fluidic configuration for LC-MS (UPLC-LCT Premier) analysis. A post-column salt diversion valve (top-left corner of the LCT Premier) can be

Page 4

Figure 5. Fluidic configuration for LC-MS analysis using mass spectrometers with no integrated salt divert valve.Figure 6. Column placement in the co

Page 5

2. Plug the terminal block into terminal block I/O (terminal block I/O is located on the back panel of the external divert valve).3. Connect the oth

Page 6

VI. SUGGESTED METHOD FOR MASS ANALYSIS USING UPLC INTACT MASS ANALYSIS APPLICATION KITMass Spectrometry has become a powerful tool for therapeu

Page 7

Figure 8. Total ion chromatograms (TICs) from Waters ACQUITY UPLC-LCT Premier ESI-TOF MS analyses of an intact IgG1, and pre and post blank runs. Col

Page 8

MS ConditionsThe MS parameters listed below can vary depending upon the physical characteristics of the analyzed proteins, flow rate, and MS probe pos