Dye Quality Control Reference Material Kit 1
[ CARE AND USE MANUAL ]
CONTENTS
I. INTRODUCTION
a. MS-Directed Purification
b. The Purpose of a QCRM
c. What Affects Your QCRM Results
II. STORAGE AND STABILITY
III. RECOMMENDED USAGE OF THE DYE QCRM
a. Testing System Performance
b. Performing the Delay Timing Test Manually
c. Performing a Chromatography Test
IV. TROUBLESHOOTING
V. ORDERING INFORMATION
DYE QUALITY CONTROL REFERENCE MATERIAL KIT
I. INTRODUCTION
The Quality Control Reference Materials (QCRM) portfolio is a unique
collection of standards and mixtures. These products allow users
to evaluate and benchmark the chromatography system before
analysis of critical material. The products in the portfolio are all
precisely formulated based on the expertise of Waters scientist.
The AutoPurification
Dye QCRM contains 3 compounds including
Thionin, Thioflavin T, and Crystal Violet.
This standard mix is used to confirm the benchmark performance
of a Preparative/Purification System through fraction collection.
This particular QCRM is precisely formulated to allow the user to
verify correct fraction collection. This test is highly valuable to
perform upon first use/installation, and any time the system
has not been used or has had any changes. These compounds were
vigorously tested and evaluated and chosen because they provide
the following advantages:
• Well-separated
• Easily visually identified
a. MS-Directed Purification
The Waters MS-directed autopurification system uses the Waters
SQD2 Mass Detector as its primary detector.
UV-Directed SystemMS-Directed System
PC with MassLynx and
FractionLynx software
2767 sample
manager
2545 binary
gradient module
Bottle
tray
2489 UV
detector
System fluidics
organizer
PC with MassLynx and
FractionLynx software
2767 sample
manager
2545 binary
gradient module
515 makeup pump
(not shown)
Bottle
tray
2998 PDA
detector
SQD2
detector
System fluidics
organizer
Page view 0
1 2 3 4 5 6 7 8

Summary of Contents

Page 1 - [ CARE AND USE MANUAL ]

Dye Quality Control Reference Material Kit 1[ CARE AND USE MANUAL ]CONTENTSI. INTRODUCTION a. MS-Directed Purification b. The Purpose of a QCRM

Page 2

Dye Quality Control Reference Material Kit 2[ CARE AND USE MANUAL ]b. The Purpose of a Quality Control Reference MaterialWaters recommends to benchmar

Page 3

Dye Quality Control Reference Material Kit 3[ CARE AND USE MANUAL ]Note: When using a different size and type of column than the one described here, t

Page 4

Dye Quality Control Reference Material Kit 4[ CARE AND USE MANUAL ] Example: If the dye is seen in the waste line 15 seconds before the valve clicks

Page 5

Dye Quality Control Reference Material Kit 5[ CARE AND USE MANUAL ]• For a UV-directed system, enter 418 in the Wavelength A column and 590 in the Wa

Page 6

Dye Quality Control Reference Material Kit 6[ CARE AND USE MANUAL ]1. From the shortcut bar, click FractionLynx > Collection Control, and verify t

Page 7

Dye Quality Control Reference Material Kit 7[ CARE AND USE MANUAL ]V. ORDERING INFORMATIONTo order these products, contact your nearest subsidiary, or

Page 8

Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com8[ CARE AND USE MANUAL ]©2013 Waters Cor

Comments to this Manuals

No comments